TECHNICAL SPECIFICATIONS
| Light Source | 6V, 10W tungsten halogen lamp, it can work over 5000 hours |
|---|---|
| Wavelength Range | 400-750 nm |
| Optical Fitters | 4 Standard filters: 405, 450, 492, 630 nm, maximum 8 filters |
| Measurement Methods | Single wavelength, dual wavelength, two points method, dynamic assay, multiple wavelengths |
| Blank Ways | Single well blank, multiple wells blank, rows and columns blank |
| Calculation Method | Absorbance, single point callibration, linear regression, quadratic curve, four parameters equation |
| Measurement Speed | Single wavelength <3 seconds/96 wells, dual wavelength <9 seconds/96 wells |
| Measurement Range | 0.0000-4.5000Abs |
| Resolution | 0.0001A |
| Linearity | ± 05% or< 0.025A |
| Repeatability Difference | < 0.1% or ± 0.0025A |
| Rows and Columns Difference | < 0.01A |
| Display | Built-in XP system, 8" touch-screen, shows whole microplate results and calibration curve |
| Save Information | More than 300 test programs, results of 50 pieces of 96 well, 100000 test results |
| Interface | 4 USB two-way communication interface, 1 RS232, 1 VGA, 1 Keyboard connector, 1 mouse connector |
| Environment | Power supply 12V DC (100V-240V wide input voltage range) temperature: 5°-40°C |
| External Dimension | 52.5cm(20.7 in) x 42.0cm(16.5 in) x 26.0cm(10.20 in) |
| Net Weight | 9 kg. |