TECHNICAL SPECIFICATIONS
- Wavelength Range :- 200-800nm
- Minimum Sample Size :- 0.5μ-2μl (2μl advised)
- Path Length :- 0.2mm or 1mm
- Light Source / Life :- Xenon flash lamp / > 109 flashes
- Detector Type :- 3864 element linear silicon CCD array
- Wavelength Accuracy :- +1 nm (TWHM at Hg 253.7 nm)
- Spectral Resolution :- ≤3nm (FWHM @ Hg 253,7 nm)
- Absorbance Precision :- 0.003Abs (1mm path length)
- Absorbance Accuracy :- <1% (7.332 Abs, at 260 nm wavelength)
- Absorbance Range :- 0.04-90 (at 260 wavelength, 10mm equivalent)
- Detection Concentration Range :- 2ng/ul dsDNA ~ 4,500ng/ul dsDNA
- Detection Time :- <8 Seconds
- Input Voltage :- DC, 24V, 2A
- Power :- 20W
- Dimension (WxDxH) :- 200 x 250 x 166mm