Description

M1180 Is Supplied Complete With:

  • MA9178/1 pH Electrode.
  • MA814DB/1 EC/TDS/NACI/Temperature probe.
  • MA831R Temperature Probe.
  • MA9315 Electrode Holder.
  • M10004 pH 4.01 Sachet Buffer Solution.
  • M10007 pH 7.01 Sachet Buffer Solution.
  • M10010 pH 10.01 Sachet Buffer Solution.
  • M10016 Sachet Electrode Cleaning Solution.
  • M10031 1413 µS/cm Sachet calibration solution.
  • Mi5200 Application Software.
  • MAQ9350 RS232 connection cable with 2 meters cable.
  • 12 VDC Adapter & Instruction manual.

TECHNICAL SPECIFICATION

pH Range -2.00 to 16.00 pH; -2.000 to 16.000 pH Rel mV Offset +2000 mV
mV Range ±699.9 mV; 2000 mV pH Calibration 1, 2 or 3 points calibration, with 7 memorized buffers (pH 1.68, 4.01. 6.86, 7.01, 9.18, 10.01 and 12.45)
EC Range 0.00 to 29.99 µS/cm; 30.0 to 299.9 LS/cm; 300 to 2999 µS/cm: 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm: up to 500.0 mS/cm uncompensed EC* EC Calibration 1 point slope calibration with 6 memorized solutions: (84 uS/cm, 1413 US/cm, 5.00 mS/cm, 12.88 uS/cm, 80.0 uS/cm, 111.8ms / cm )
TDS Range 0.0 to 14.99 mg/l (ppm): 15.0 to 149.9 mg/l (ppm): 150 to 1499 mg/l (ppm): 1.50 to 14,99 g/l (ppt); 15.0 to 100.0 g/l (ppt); up to 400.0 g/l actual TDS* (with 0.80 factor) NaCI Calibration Temp Calibration Temperature Compensation 1 point, with MAS8066 solution 2 point, at 0 and 50 deg * C / 32 and 122 deg * F automatic or manual, from -20.0 to 120.0°C/-4.0 to 248 deg * FF selectable from 0.00 to 6.0096 /^ C EC and TDS only)
NaCl Range 0.0 to 400.0%
Temp Range -20.0 to 120.0°C/-4.0 to 248.0°F Temperature Coefficient
PH Resolution 0.01 pH: 0.001 pH
mV Resolution 0.1 mV; 1 mV pH Electrodes & Temp Probe MA814DB/1 & MA917B/1 & MA831R (included)
EC Resolution 0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm; TDS Factor 0.40 to 0.80 (default value is 0.50)
TDS Resolution 0.01 mg/i; 0.1 mg/l; 1.0 mg/l; 0.01 g/l; 0.1 g/l Log on demand up to 50 samples on each range (pH, mV, EC, TDS, NaCl)
NaCl Resolution 0.1%
Temp Resolution 0.1°C/0.1°F GLP last pH, EC, NaCI calibration data
PH Accuracy ±0.01 pH; ±0.002 pH PC Interface RS232/USB Opto – isolated
MV Accuracy ±0.2 mV; 1 mV Environment 0 to 50°C/32 to 122°F; max RH 95%
EC Accuracy ±1% of reading ±(0.05 pS/cm or 1 digit) Power supply 12 VDC power adapter (included)
TDS Accuracy ±1% of reading ±(0.03 ppm or 1 digit) Dimensions 230 x 160 x 95 mm
NaCl Accuracy ±1% of reading Weight 0.9 kg
Temp. Accuracy ±0.4°C/+0.8°F