Description
Construction:
Robust construction and enhanced safety features qualify the cabinet for the most demanding laboratory Applications. The cabinet is fully assembled and ready to install.
- Main Body Made of 18 gauge electro-galvanized steel, with oven backed epoxy powder coated finish.
- The interior work area is formed from a single pieces of stainless steel 304 grade with large radius corners to simplified cleaning.
- The cabinet working zone has no welded joints to collect contamination and rust .
- The working table made of S.S 304 Grade.
- The interior work area is formed from a single pieces of stainless steel 304 grade with large radius corners to simplified cleaning.
- The cabinet working zone has no welded joints to collect contamination and rust .
- The working table made of S.S 304 Grade.
- Front door has a unique feature of sliding up and down with very light hand pressure as counter balance with another weight.
- The lower drain through the single piece fabrication with wide open angle and channel to direct spills to the drain. .
- There is no screws in the front or sides to trap contaminants or complicate cleaning.
- Electrical outlets are mounted below service fixtures to minimize obstruction
Key Benefit of using the EC motor:
- Monitoring and fine adjustment of operating modes with the microprocessor control system
- Low energy consumption.
- Low heat emission.
- Low noise level.
- No vibration.
- Longer operating life.
Soft Touch Keyboard Membrane:
Soft Touch Keyboard Membrane touch keyboard and the rear-lit LCD display all required data keeping the operator constantly up to date of the cabinet conditions
- Estimated life of HEPA filters
- Estimated life of UV lamp
- Front sash position.
- Air Velocity: Inflow & Down Flow .
- Both Temperature.
- Two Mode: Quick & Normal.
- Front sash position.
Microprocessor control & Alarms System:
- The microprocessor controller shall be mounted on the main control panel ,Angled down toward the user for easy access.
- The controller shall include soft touch key pad controls and backlit LCD display to access operation of the blower /motor , Light ,UV Light and Menu.
- On the LCD display simultaneously display mode no, Sash Status, Real clock, Inflow, Down Flow, Air flow status and temperature.
- Audible and Visual alarms shall be provided for Unsafe conditions such as improper airflow or Sash.
- The air flow shall be monitored by air velocity sensor mounted in the cabinet.
- The cabinet shall have field calibration mode that simplifies on site calibration mode.
- In this cabinet there is two type of operational mode -Normal mode secondary Quick start mode.
- A Selectable Quick start mode should be available to automatically turn the blower and light ON/OFF By moving Sash window to the correct position.
- Airflow sensor alerts the user if airflow is sufficient for greater safety.
- Password protection (Programmable) for authorized operation.
- Programmable UV Light to allow specific exposure time from 0 to 60 minute with automatic shut-off features UV interlocking with front window i.e. closed automatically when front window is open.
- UV timer showing used time and balance available life in hours.
- Total HEPA Filter timer showing used time and balance available life in hour.
- UV light/Door/Fluorescent Light Inter Lock i.e. UV switches ON only when door is closed, and automatically switches OFF fluorescent lights. Display of safe / unsafe position for door
Quest Class II B2 Biological Safety cabinet:
- Featuring 100% exhaust, protects lab staff, the environment and sensitive work processes in which biological agents are applied.
- The cabinet offers a high level of contamination protection, based on two advanced HEPA filters operating at an efficiency of @99.99% @ 0.3 um.
- The cabinet is equipped with a smart, safe and elegant touch-screen control system that protects the operator and provides alerts for periodic maintenance actions and devices’ replacement.
Special Features:
1. Password protection (Programmable) for authorized operation.
2. Air velocity Control through the microprocessor.
3. Two Type Of Mode: Normal Mode, Quick Mode.
4. 30% enhance life of HEPA filter due to speed controller.
5. Real time Cock.
6. Microprocessor based controller.
7. UV timer showing used time and balance available life in hours
8. Total HEPA Filter timer showing used time and balance available life in hour
9. HEPA filter alarm after HEPA filter life is over and guideline for further action through LCD controller.
10. UV light/Door/Fluorescent Light Inter Lock i.e. UV switches ON only when door is closed, and switches OFF fluorescent lights.