Description

TECHNICAL SPECIFICATIONS

Throughput 140 test per hour
Test Methods End point Kinetic, Fixed time, Bi-chromatic, Multi-standard, Two-step end-point and ect
Calibration Linear or Nonlinear with 1-6standards calibration
System Calibration Deduction of water blank / reagent blank
Sample Position 64 positions including sample, calibrabtor/standard, control and stat
Sample Volume 1~100µl, 1µl/step
Sample Probe With liquid detection and crash-protect function
Reaction Tray 96 reaction cuvettes
Flow Cell 32ul, automatically washing
Optical System 6V/10W halogen lamp 7 high quality narrow band filters and 2 more filter positions Photometric measuring range: 0-3.0 0.0.
Diluter 1000µl 1.0µl/step
Temperature 37°C +0,1°C
Storage Capacity Memory of more than 200 programs and one million test results
External Devices (on request) Premium IV PVC, printer, monitor
Power Supply AC 110/220V~+/-10%, 50/60Hz +/-1Hz
Interface USB 2.0
Dimension 60cm X 50cm X 41cm

SALIENT FEATURES

  • Batch analyzer
  • Insert STAT sample at any time during testing
  • Online display analyze state, giving immediate analysis information
  • Intelligent multi-task software
  • User-friendly software in English, French, Spanish, Russian versions, and other languages (software available on request)
  • Perfect quality control function with Levy-Jennings chart and Westguard analysis available
  • Automatic Flow cell washing and probe washing between each test to effectively minimize carry-over
  • Auto-check and Auto-alarm to ensure the analyzer is in good condition
  • New long-life pump, no tubing wearing problem anymore
  • Less service and maintenance needed