Description

TECHNICAL SPECIFICATIONS

General Range Serum Range Urine Range
ppm meq/1 meq/1
Na: 0-100 0-200 megq/1, 1:100 dil 0-250 megq/1, 1:100 dil.
K: 0-100 0-10 meq/l, 1:100 diL 0-100 meq/1, 1:100 dil
Ca: 0-100 0-10 mea/l, 1:2 dil.
Li: 0-100 0-2 meq/L, 1:5 dil.
Ba: 30-3000 (all meq values are claimed with appropriate dilution)
Sensitivity General Mode: Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca:15 ppm Ba: 30ppm
Filters Na & K Supplied (Ca, Li & Ba are optional at extra cost) & Factory Fitted only
Resolution 0.1 ppm/0.1 meq.
Reproducibility < 1% coefficient of variation for 20 consecutive samples
using 10 ppm Na set as Maximum Standard up to 3% CV for 20 consecutive samples of Concentrations between 50 to 199 ppm for Barium.
Linearity < 2% error when 3 ppm Na/K and 5 ppm Li are set as Maximum Std.
Curve Fit Accuracy ± 2% of FS
Display 128 x 64 Graphical LCD
Average Time In-built in Software
Flame System LPG & Dry Oil Free Air
Detector Photodiode
Calibration With up to 25 Standards of user choice with curve fitting software (5 for each Element). Multiple Calibration Curves can be stored in the memory
Aspiration Time 3 ml per Minute
Gas Control Adjustable with knob
Flame Failure Auto Detection
Gas Cuttoff Automatic
Ignition System Auto Ignition
Automiser Axial Flow Type
Dimensions Instrument: 365 x 245 x 220 mm Approx. (LxBxH)
Pc Interface USB
Weight Main Unit: 8 Kg. (Approx.)
Power Supply 230 V AC ± 10%, 50 Hz

SALIENT FEATURES

  • Auto Gas Cutoff Facility
  • Auto Flame Failure Detection Facility
  • Inbuilt Thermal Printer (In Model QST-0553)
  • UP Based With Printer Interface
  • Five Element Measurement In Single Aspiration of the Sample
  • Curve Calibration Programmability using Maximum of 25 Standards
  • Data Processing Through Curve Fitting Techniques -Least Squares, Quadratic
  • Multiple  Calibration Curves Saving Facility
  • Samples Data Storage Facility
  • Online Help Key
  • Printer Attachment Facility
  • Measurement in ppm & meq Useful For Both Medical & Soil Testing Applications
  • Setup Storage Facility up to 10 setups files
  • Supplied with Compressor Unit

STANDARD ACCESSORIE

  • Air Compressor
  • Chimney
  • Sample Beakers
  • Manual
  • Air & Gas Tubing
  • USB Cable
  • Main Lead
Related